Contact person
Louise Wogelred
Forsknings- och utvecklingsingenjör
Contact LouiseIn the Instruction for Users (IFU), supplied with reusable medical devices or medical devices to be sterilized by the user, there should be a validated instruction for sterilization. We can validate the sterilization of your product, according to EN ISO 17665:2024.
Contribute to healthcare safety by providing clear and validated instructions for sterilization.
Fulfill requirements in MDR.
Sterilization validation is performed by showing a sterility assurance level (SAL) of 10-6 using, for the method most resistant organism, Geobacillus stearothermophilus, at a concentration enough to show a sterility assurance level (SAL) of minimum 10-6. Inoculated devices are packed according to the instructions and processed with moist heat following the partial cycle approach i.e. a shorter time than stated in the instructions is used. Potential surviving spores on the devices are incubated in growth medium and evaluated for turbidity according to ISO 11737-2.
We can also evaluate that the drying time after sterilisation is enough, to ensure safe storage of the product. RISE provides a Systec DX-90, validated according to ISO 17665 with external temperature probes.
We can also validate the instructions for cleaning and disinfection according to ISO 17664-1/2, ISO 15883-5 and ANSI/AAMI ST98: Validation of instructions and processes for cleaning and disinfection of medical devices | RISE
Methods and results are presented in a written report.